Oxygen Index Tester

Oxygen Index Tester

Product Description

The Oxygen Index Tester is used to determine the minimum oxygen concentration (oxygen index) required to just sustain combustion of materials in a combustion cylinder with a controlled mixture of oxygen and nitrogen. This value is used to evaluate the combustion performance classification of materials or for product quality control.

Scope of Application

This instrument is suitable for testing the combustion performance of strip- or block-shaped materials with a thickness less than 10.5 mm that can support themselves vertically. It also applies to homogeneous solid materials, laminated materials, or foam materials with an apparent density greater than 100 kg/m³. Additionally, it is suitable for certain foam materials with an apparent density less than 100 kg/m³.
Typical materials include plastics, rubber, insulation materials, flame-retardant fabrics, composite materials, extruded boards, fiberglass-reinforced plastics, wires and cables, waterproofing materials, fibers, ropes, flame-resistant materials, foams, polyurethane, asphalt, soft films, wood, films, etc.

Note: For evaluating flame spread characteristics of foam materials with a density below 100 kg/m³, relevant equipment specified in GB/T 8332 should be used.

Applicable Standards

Conforms to the following standards:

GB/T 2406.2-2009

ISO 4589.2-1996

GB 8624-2012

Technical Specifications

  1. Equipment Specifications

Oxygen concentration measurement range: 0–100%

Measurement accuracy: ±0.1%

Display mode: Digital display

Zero adjustment: Available

Full-scale adjustment: Available

Working pressure: 0.05–0.15 MPa

Pressure gauge range: 0–0.4 MPa

Pressure gauge accuracy: Grade 2.5

Flow rate adjustment range: 0–15 L/min

Flow meter accuracy: Grade 2.5

Response time: <5 seconds

Combustion cylinder inner diameter: 75 mm

Combustion cylinder height: 480 mm (excluding taper inlet)

Gas flow velocity in combustion cylinder: 40 mm ±2 mm/s

  1. Operating Environment & Gas Supply

Environment: Temperature 23°C ±2°C, relative humidity ≤70%

Power supply: AC 220V ±5%, 50 Hz

Power consumption: 50 W

Gases used: Oxygen and nitrogen with purity ≥98% (gas cylinders, pressure regulators, and dryers are provided by the user)

Gas supply pressure: 0.1–0.3 MPa